To Study variations in Blood Pressure (BP) Recording by Mercurial Sphygmomanometer & by LED (Non-Mercurial) BP Machine in Individuals in Sitting Position in Same Arm - A Cross Sectional Study in Guwahati City
International Journal of Science and Research (IJSR)

International Journal of Science and Research (IJSR)
Call for Papers | Fully Refereed | Open Access | Double Blind Peer Reviewed

ISSN: 2319-7064


Downloads: 3 | Views: 254 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1

Research Paper | Physiology | India | Volume 10 Issue 5, May 2021 | Popularity: 5.2 / 10


     

To Study variations in Blood Pressure (BP) Recording by Mercurial Sphygmomanometer & by LED (Non-Mercurial) BP Machine in Individuals in Sitting Position in Same Arm - A Cross Sectional Study in Guwahati City

Arijit Mazumdar, Reeta Baishya


Abstract: Background: Arterial blood pressure ( BP ) is an important component of our body parameter & its maintenance to near normal levels is important for maintenance of internal milieu of our body .Mercurial BP instruments are in the process of being phased out due to environmental hazards posed by the liquid metal. Methods: After obtaining ethical clearance from institutional ethical committee a total of 100 individuals between 20-60 yrs of age of either gender were taken in the ambit of the study. The BP of the individuals were measured in the same arm in sitting position by Mercury Sphygmomanometer & by LED BP recording machine at room temperature after relaxing the subjects and readings were recorded accordingly in a profoma made beforehand . Prior to BP being taken the individuals were asked to sign the written consent & a brief history & physical examination was done. The BP readings with both Mercurial sphygmomanometer & LED instrument were tabulated & variation of readings & their statistical significance was found out For statistical analysis the value of the BP recordings were presented as Mean? standard deviation. Statistical correlation between the recordings of BP by both methods were done and P-value<0.05 was considered as significant. Results: BP readings recorded with both the instruments were found to be similar( readings of Mercurial & LED BP instrument) as result showed non significant ?p? value 0.1001 (systolic) & 0.1180(diastolic) (p>0.05) . The mean of the arterial pressure as recorded by mercurial & LED were similar. Conclusions: It was therefore concluded from the study, that recording of BP by Mercurial & LED instruments yielded almost similar readings. So, by phasing out mercurial BP instruments, LED instruments are a good option for measuring arterial BP & can be used.


Keywords: LED BP Instrument, Mercurial Sphygmomanometer, Hypertension


Edition: Volume 10 Issue 5, May 2021


Pages: 1113 - 1115


DOI: https://www.doi.org/10.21275/SR21524214801


Please Disable the Pop-Up Blocker of Web Browser

Verification Code will appear in 2 Seconds ... Wait



Text copied to Clipboard!
Arijit Mazumdar, Reeta Baishya, "To Study variations in Blood Pressure (BP) Recording by Mercurial Sphygmomanometer & by LED (Non-Mercurial) BP Machine in Individuals in Sitting Position in Same Arm - A Cross Sectional Study in Guwahati City", International Journal of Science and Research (IJSR), Volume 10 Issue 5, May 2021, pp. 1113-1115, https://www.ijsr.net/getabstract.php?paperid=SR21524214801, DOI: https://www.doi.org/10.21275/SR21524214801