Downloads: 1 | Views: 97 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Research Paper | Medical Science | India | Volume 12 Issue 3, March 2023
Development of Test Method for Assessment of In-Vitro Degradation of Medical Devices
Minocha Dr. Pramod Kumar [2] | Kothwala Dr. Deveshkumar Mahendralal [2] | Shaikh Amirhamzah Mahmadiqbal [2] | Patel Chirag P
Abstract: The objective of this study is to present test results that have been reported from an in-vitro, real-time degradation study of a circular knitted mesh of a biodegradable medical device. The knitted mesh on the stent eroded up until that point in the trial, which lasted for 6-7 months at 37??1?C. Hence, in this study, the degradation of a circular knitted mesh formed of 28 micron monofilament Poly D, L-lactide-co-glycolic acid (PLGA) 8523 of biodegradable medical device was evaluated under in-vitro, real-time degradation conditions.
Keywords: Knitted mesh, biodegradable, in-vitro degradation, real time
Edition: Volume 12 Issue 3, March 2023,
Pages: 1553 - 1557