Downloads: 145 | Weekly Hits: ⮙1 | Monthly Hits: ⮙2
Case Studies | Hematology | India | Volume 6 Issue 1, January 2017
Spurious High Platelet Count by Automated Hematology Analyzer in Patient with Microcytic RBCS
Dr Rateesh Sareen [2] | Dr Menka Kapil
Abstract: We report a case of spurious elevation of platelet count on three consecutive occasions in a pediatric patient on six part Sysmex Hematology auto analyzer. Falsely low platelet counts are frequently reported in literature but falsely elevated Platelet counts are uncommon. The case emphasizes the important role of careful peripheral blood smear examination in case of high platelet counts. The blind folded reliance on automatic analyzers in hematology should be discouraged so as to provide best patient care.
Keywords: Spurious platelet count, 5 part analyzer
Edition: Volume 6 Issue 1, January 2017,
Pages: 526 - 527
Similar Articles with Keyword 'platelet'
Downloads: 1 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Review Papers, Hematology, India, Volume 12 Issue 3, March 2023
Pages: 1386 - 1393Immature Platelet Fraction: A Promising Biomarker in Various Clinical Entities
Kavitha Thilak [3] | Abin Varghese [2] | Vineetha V [6]
Downloads: 3 | Weekly Hits: ⮙2 | Monthly Hits: ⮙2
Case Studies, Hematology, India, Volume 12 Issue 5, May 2023
Pages: 308 - 310Evans Syndrome: A Case Report
Motij Dalai | Deepak S [23] | Emmanuel Paul [2]